Simple, stable and sensitive electrogenerated chemiluminescence detector for high-performance liquid chromatography and its application in direct determination of multiple fluoroquinolone residues in milk
Authors:Li Yongbo  Zhang Zhujun  Li Jinsong  Li Hongguang  Chen Yan  Liu Zhaohui
Institution:a Key Laboratory of Analytical Chemistry for Life Science of Shaanxi Province, School of Chemistry and Materials Science, Shaanxi Normal University, Xi’an 710062, China
b Xi’an Center for Disease Control and Prevention, Xi’an 710054, China
Abstract:A simple, stable and sensitive electrogenerated chemiluminescence (ECL) detector was developed. It was based on tris(2,2-bipyridyl)ruthenium(II) (Ru(bpy)32+) immobilized on the surface of a Pt wire with Nepem-105D ion exchange solution. The detector was prepared by inserting a Pt wire with immobilized Ru(bpy)32+ (working electrode) into a capillary tube, followed by inserting another Pt wire (counter electrode) in this tube and sealing. ECL behavior was investigated using ofloxacin as an analyte. Under optimal conditions, stable ECL intensity was obtained. This detector has been used in HPLC-ECL for the determination of multiple target fluoroquinolone residues in milk. There is no post column reagent addition, which would dilute the analytes, potentially leading to chromatographic band-broadening. The system is very simple with low dead volume, low baseline and background noise, together with high sensitivity and stability. The as-prepared ECL detector, when was used for the determination of ofloxacin, pefloxacin, enrofloxacin and difloxacin in milk, demonstrated adequate sensitivity to allow quantification of trace FQ levels in commercial milk samples. One or more of the target FQ analytes were present at levels above the LOD of the new ECL detector in each and every one of the 22 milk samples analysed.
Keywords:Electrogenerated chemiluminescence detector  Tris(2  2-bipyridyl)ruthenium(II)  High performance liquid chromatography  Multiple fluoroquinolone residues
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号