Abstract: A new method for the direct determination of lead traces using derivative atom trapping flame atomic absorption spectrometry (DAT-FAAS) with an improved water-cooled stainless steel trapping equipment in an air-acetylene flame was investigated. The optimum conditions concerning the sensitivity were studied. For a 1 min collection, the characteristic concentration (given as derivative absorbance of 0.0044) and the detection limit (3s) were 1.4 ng/mL and 0.27 ng/mL, respectively. This is 361 and 74-fold better than those of the conventional flame atomic absorption spectrometry (FAAS) and comparable to those of graphite furnace atomic absorption spectrometry (GFAAS). The detection limit and sensitivity of DAT-FAAS for a 3 min collection time were 2 and 3 orders of magnitude higher than those of conventional FAAS. The present method was applied to the determination of lead in water and liqueur samples with a recovery range of 94–108% and a relative standard deviation of 3.5–5.6%. Received: 10 January 1996/Revised: 9 December 1996/Accepted: 20 December 1996