摘    要:采用高纯氧化物经四硼酸锂和偏硼酸锂熔融制备人工标准样品,以氧化镱为内标,制作校准曲线,建立了波长色散型X射线荧光光谱法测定钴精矿中钴、铜、锰元素的分析方法。重点研究了混合熔剂、试样稀释比、氧化剂和内标选择、方法检出限、方法准确度和精密度等,结果表明,各元素校准曲线线性范围宽,相关系数均大于0.999, 钴、铜、锰元素检出限分别达到0.002%、0.001%和0.001%,测试准确度和精密度可靠,方法满足行业检测需求。

关 键 词:X射线荧光光谱法  钴精矿        熔融制样
收稿时间:2021/1/15 0:00:00
修稿时间:2021/6/4 0:00:00

Determination of cobalt, copper and manganese in cobalt concentrate by wavelength dispersive X-ray fluorescence spectrometry
FENG YA HUI,HOU JIAN JUN,PAN Shenglin,DAI Dongqing,ZHANG XIU,WANG Xuning,XU REN FU and MENG QIN SEN.Determination of cobalt, copper and manganese in cobalt concentrate by wavelength dispersive X-ray fluorescence spectrometry[J].Chinese Journal of Inorganic Analytical Chemistry,2021,11(4):13-18.
Authors:FENG YA HUI  HOU JIAN JUN  PAN Shenglin  DAI Dongqing  ZHANG XIU  WANG Xuning  XU REN FU and MENG QIN SEN
Institution:Nanjing Customs Industrial Products Testing Center,Nanjing Customs Industrial Products Testing Center,Nanjing Customs Industrial Products Testing Center,Nanjing Customs Industrial Products Testing Center,Nanjing Customs Industrial Products Testing Center,Nanjing Jinli Inspection Limited Company,,Nanjing Customs Industrial Products Testing Center,Nanjing Jinli Inspection Limited Company
Abstract:A method for the determination of cobalt, copper and manganese in cobalt concentrate by wavelength dispersive x-ray fluorescence spectrometry was established. The artificial standard sample was prepared by melting high-purity oxides with lithium tetraborate and lithium metaborate. The calibration curve was made with ytterbium oxide as internal standard. The results show that the calibration curve of each element has a wide linear range, the correlation coefficient is greater than 0.999, and the detection limits of cobalt, copper and manganese are 0.002%, 0.001% and 0.001% respectively. The test accuracy and precision are reliable, and the method meets the needs of the industry.
Keywords:x-ray fluorescence spectrometry  cobalt concentrate  cobalt  copper  manganese  fusion sample preparation