首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of Cysteine at Bismuth Nanostructure – Carbon Ionic Liquid Electrode by Square Wave Voltammetry
Authors:Afsaneh Safavi  Abdolkarim Abbaspour  Maral Ajamian
Institution:Department of Chemistry, College of Sciences, Shiraz University, Shiraz 71454, Iran
Abstract:Bismuth nanostructure‐carbon ionic liquid electrode has been employed for sensitive determination of cysteine (Cys). Bismuth nanostructure was incorporated into the carbon ionic liquid electrode (CILE) and applied for determination of cysteine. An interaction was taking place between bismuth nanostructure and the thiol group of cysteine. The bismuth cysteinate complex oxidation potential occurred at more negative potential compared to the cysteine oxidation peak obtained at bare carbon ionic liquid electrode (CILE). Square wave voltammetry (SWV) was used for the determination of the cysteine and satisfactory results were obtained. The calibration curve was linear in the concentration range of 1 to 500 µM and 0.5–2 mM of Cys. A low detection limit of 0.5 µM was achieved for Cys. The electrode showed a good selectivity for determination of cysteine in the presence of other amino acids. The proposed electrode was satisfactory applied for the determination of cysteine in human serum plasma sample.
Keywords:Cysteine  Bismuth nanostructure  Carbon ionic liquid electrode  Square wave voltammetry
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号