首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Photoelectrochemical‐assisted Batch Injection Analysis (PEC‐BIA) of Glucose Exploiting Visible LED Light as an Excitation Source
Authors:Leonardo Carvalho Soares  Clenilton Costa dos Santos  Rita de Cssia Silva Luz  Flvio Santos Damos
Institution:Leonardo Carvalho Soares,Clenilton Costa dos Santos,Rita de Cássia Silva Luz,Flávio Santos Damos
Abstract:This work describes the development of a novel method for glucose determination exploiting a photoelectrochemical‐assisted batch injection analysis cell designed and constructed with the aid of 3D printer technology. The PEC‐BIA cell was coupled to a LED lamp in order to control the incidence of light on the Cu2O/Ni(OH)2/FTO photoelectroactive platform. The electrochemical characteristics of Cu2O/Ni(OH)2/FTO photoelectroactive platform were evaluated by cyclic voltammetry, amperometry, and electrochemical impedance spectroscopy. The PEC‐BIA cell presented linear response range, limit of detection based on a signal‐to‐noise ratio of three, and sensitivity of 1–1000 μmol L?1, 0.76 μmol L?1 and 0.578 μA L μmol?1, respectively. The PEC‐BIA method presented a mean value of the recovery values of 97.0 % to 102.0 % when it was applied to glucose determination in artificial blood plasma samples which indicates the promising performance of the proposed system to determine glucose.
Keywords:Photoelectrochemistry  Cu2O/Ni(OH)2  3D-Printing  Batch injection analysis  Glucose  Visible LED light
设为首页 | 免责声明 | 关于勤云 | 加入收藏