Electrochemical Immunoanalysis for Carcinoembryonic Antigen Based on Multilayer Architectures of Gold Nanoparticles and Polycation Biomimetic Interface on Glassy Carbon Electrode
Abstract:This paper describes a layer‐by‐layer (LBL) self‐assembly process of chitosan (CTS) and gold nanoparticles (Au) on the pretreated negatively charged glassy carbon (GC) electrode to fabricate electrochemistry immunosensor with a nontoxic biomimetic interface, which provided an environment similar to a native system and allowed more freedom in orientation for immobilization of carcinoembryonic antibody (anti‐CEA) to monitor carcinoembryonic antigen (CEA). UV‐vis spectroscope, atomic force microscopy (AFM), and cyclic voltammetric (CV) measurements were used to follow the multilayer film formation. The performance of the biominetic interface and factors influencing the assay system were investigated in detail. The differential pulse voltammetry (DPV) current response is used for the CEA concentration assay. The dynamic range was from 0.50 to 80.00 ng mL?1 with a detection limit of 0.27 ng mL?1 at 3σ. In addition, the experiment results indicate that immobilization described in this proposed method exhibits a good sensitivity, selectivity, and stability.
Keywords:Biomimetic interface selection  Layer‐by‐layer assembly  Gold nanoparticles  Chitosan  Immunoanalysis