Polydivinylbenzene/Ethylvinylbenzene Composite Membranes for the Optimization of a Whole Blood Glucose Sensor
Abstract:A novel ultra thin polydivinylbenzene/ethylvinylbenzene composite membrane has been developed for use as the outer covering barrier in a model amperometric glucose oxidase enzyme electrode. The composite membrane was formed via the cathodic electropolymerization of divinylbenzene/ethylvinylbenzene at the surface of gold sputter coated host alumina membranes, (serving solely as a mechanical support for the thin polymer film). Permeability coefficients were determined for the enzyme substrates, O2 and glucose, across composite membranes formed with a range of polymer thicknesses. Due to the highly substrate diffusion limiting nature of the composite membrane, it was found that anionic interferents present in blood (such as ascorbate), were effectively screened from the working electrode via a charge exclusion mechanism, in a manner similar to previous findings within our laboratory. The enzyme electrode showed an initial 32% signal drift when first exposed to whole human blood over a period of 2 hours, after which time enzyme electrode responses remained essentially stable. Whole blood patient glucose determinations yielded a correlation coefficient of r2=0.97 in comparison to standard hospital analyses.
Keywords:Cathodic electropolymerization  Composite membrane  Amperometric enzyme electrode     Polydivinylbenzene  Blood