Atomic distributions in acetylene-air flame using hydraulic high-pressure nebulization with direct (100%) introduction of solutions Applications to speciation analysis |
| |
Authors: | Áron Béni Sándor Berényi |
| |
Institution: | a Department of Inorganic and Analytical Chemistry, University of Debrecen, Egyetem tér 1, H-4032, Debrecen, Hungary b Department of Agricultural Chemistry, University of Debrecen, Böszörményi út 138, H-4032, Debrecen, Hungary |
| |
Abstract: | One of the fundamental factors of the performance of atomic spectrometric methods is the efficiency of sample introduction into flame or plasma. By utilization of the formerly developed hydraulic high-pressure nebulizer, a nebulizer/burner system with solution uptake efficiency of 100% have been worked out. Optimizing operation parameters of nebulizer/burner system, an improvement of 2-12 times was achieved for the signal-to-noise ratio in the flame emission and flame atomic absorption spectrometry, depending on sample flow rate. This nebulizer is especially suitable for development of liquid chromatography-AAS/FES coupled technique for speciation analytical tasks. |
| |
Keywords: | Flame atomic absorption Sample introduction Hydraulic high-pressure nebulization Speciation |
本文献已被 ScienceDirect 等数据库收录! |
|