首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Mass spectrometry monitoring of sevoflurane in the breathing circuit of an inhalation anesthesia machine
Authors:V A Elokhin  T D Ershov  A I Levshankov  V I Nikolaev  M F Saifullin  A Yu Elizarov
Institution:1.ZAO Nauchnye Pribory,St. Petersburg,Russia;2.Kirov Military Medical Academy,St. Petersburg,Russia;3.Ioffe Physical Technical Institute,Russian Academy of Sciences,St. Petersburg,Russia
Abstract:The results of controlling the amount of inhalation anesthetics sevoflurane in the patient breathing circuit of an inhalation anesthesia machine (IAM) by mass spectrometry are presented. A vacuum system with a differential chamber providing pressure difference in the range 1 × 105–1.5 ×10−4 Pa was used to inject the studied gas sample from the delivery circuit into the mass spectrometer. The concentrations of the anesthetic obtained using mass and IR spectrometry are compared. The potency of mass spectrometry for monitoring the anesthetic gas in the real time mode is demonstrated. The time dependences of the concentration of the anesthetic gas corresponding to different periods of anesthesia are given.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏