High-speed gradient parallel liquid chromatography/tandem mass spectrometry with fully automated sample preparation for bioanalysis: 30 seconds per sample from plasma
Authors:Deng Yuzhong  Wu Jing-Tao  Lloyd Thomas L  Chi Cecilia L  Olah Timothy V  Unger Steve E
Institution:Discovery MAP Group, Bristol-Myers Squibb Company, Wilmington, DE 19880, USA. Alan.Deng@bms.com
Abstract:In this work, a high-throughput and high-performance bioanalytical system is described that is capable of extracting and analyzing 1152 plasma samples within 10 hours. A Zymark track robot system interfaced with a Tecan Genesis liquid handler was used for simultaneous solid-phase extraction of four 96-well plates in a fully automated fashion. The extracted plasma samples were injected onto four parallel monolithic columns for separation via a four-injector autosampler. The use of monolithic columns allowed for fast and well-resolved separations at a considerably higher flow rate without generating significant column backpressure. This resulted in a total chromatographic run cycle time of 2 min on each 4.6 x 100 mm column using gradient elution. The effluent from the four columns was directed to a triple quadrupole mass spectrometer equipped with an indexed four-probe electrospray ionization source (Micromass MUX interface). Hence, sample extraction, separation, and detection were all performed in a four-channel parallel format that resulted in an overall throughput of about 30 s per sample from plasma. The performance of this system was evaluated by extracting and by analyzing twelve 96-well plates (1152) of human plasma samples spiked with oxazepam at different concentrations. The relative standard deviation (RSD) of analyte sensitivity (slope of calibration curve) across the four channels and across the 12 plates was 5.2 and 6.8%, respectively. An average extraction recovery of 77.6% with a RSD of 7.7% and an average matrix effect of 0.95 with a RSD of 5.2% were achieved using these generic extraction and separation conditions. The good separation efficiency provided by this system allowed for rapid method development of an assay quantifying the drug candidate and its close structural analog metabolite. The method was cross-validated with a conventional liquid chromatography/tandem mass spectrometry (LC/MS/MS) assay.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号