Multielemental analysis of dried residue from metal-bearing waters by wavelength dispersive X-ray fluorescence spectrometry |
| |
Authors: | Oscar Gonzalez-Fernandez Eva MarguiIgnacio Queralt |
| |
Institution: | Laboratory of X-ray Analytical Applications, Institute of Earth Sciences “Jaume Almera”, CSIC, Solé Sabarís, s/n 08028 Barcelona, Spain |
| |
Abstract: | The purpose of this work was evaluation of instrumental sensitivity and detection limits for determination of elemental composition (20 different elements ranging from Na to Pb) of liquid mining samples by using conventional Wavelength Dispersive X-Ray Fluorescence (WDXRF) instrumentation. Preconcentration of elements from liquid samples was performed by means of a simple dried residue process, and spectral evaluation was carried out by integration of the peak area (using WinQXAS/AXIL package software, International Atomic Energy Agency (IAEA)) instead of the common net peak line intensity traditionally used in conventional WDXRF systems. With the proposed methodology, the calculated detection limits were in the µg L− 1 range (from 0.005 to 0.1 mg L− 1 level depending on the element) in all cases, which is suitable for element determination in most liquid samples involved in environmental studies such as metal mining liquid effluents. The detection limits are also below the established limits of the TCLP 1311 (United States Environmental Protection Agency (US-EPA)) and DIN 38414-S4 (German Standard legislation) procedures, which are commonly used to evaluate the leaching of metals from landfill disposal. |
| |
Keywords: | WDXRF Liquid Sensitivity Limit of detection Mining sample |
本文献已被 ScienceDirect 等数据库收录! |
|