Iron pentacarbonyl detection limits in the cigarette smoke matrix using FT-IR spectroscopy
Authors:Parrish Milton E  Plunkett Susan E  Harward Charles N
Institution:Philip Morris USA, Research Center, P.O. Box 26583, Richmond, VA 23234, USA.
Abstract:Endogenous metals present in tobacco from agricultural practices have been purported to generate metal carbonyls in cigarette smoke. Transition metal catalysts, such as iron oxide, have been investigated for the reduction of carbon monoxide (CO) in cigarette smoke. These studies motivated the development of an analytical method to determine if iron pentacarbonyl Fe(CO)(5)] is present in mainstream smoke from cigarette models having cigarette paper made with iron oxide. An FT-IR puff-by-puff method was developed and the detection limit was determined using two primary reference spectra from different sources to estimate the amount of Fe(CO)(5) present in a high-pressure steel cylinder of CO. We do not detect Fe(CO)(5) in a single 35 mL puff from reference cigarettes or from those cigarette models having cigarette paper made with iron oxide, with a 30-ppbV limit of detection (LOD). Also, it was shown that a filter containing activated carbon would remove Fe(CO)(5).
Keywords:
本文献已被 PubMed 等数据库收录!