首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development of a sequential injection chromatography (SIC) method for determination of simazine,atrazine, and propazine
Authors:Luciana B O dos Santos  Carlos M C Infante  Jorge C Masini
Institution:Instituto de Química, Universidade de S?o Paulo, C.P. 26077, S?o Paulo, SP, Brazil. Fax: +55‐11‐3815‐5579
Abstract:This paper describes the development of a sequential injection chromatography (SIC) procedure for separation and quantification of the herbicides simazine, atrazine, and propazine exploring the low backpressure of a 2.5 cm long monolithic C18 column. The separation of the three compounds was achieved in less than 90 s with resolution >1.5 using a mobile phase composed by ACN/1.25 mmol/L acetate buffer (pH 4.5) at the volumetric ratio of 35:65 and flow rate of 40 μL/s. Detection was made at 223 nm using a flow cell with 40 mm of optical path length. The LOD was 10 μg/L for the three triazines and the quantification limits were of 30 μg/L for simazine and propazine and 40 μg/L for atrazine. The sampling frequency is 27 samples per hour, consuming 1.1 mL of ACN per analysis. The proposed methodology was applied to spiked water samples and no statistically significant differences were observed in comparison to a conventional HPLC–UV method. The major metabolites of atrazine and other herbicides did not interfere in the analysis, being eluted from the column either together with the unretained peak, or at retention times well‐resolved from the studied compounds.
Keywords:Environmental analysis  Separation methods  Triazines
设为首页 | 免责声明 | 关于勤云 | 加入收藏