Pressurized gradient capillary electrochromatographic separation of eighteen amino acid derivatives
Authors:Ru Q H  Yao J  Luo G A  Zhang Y X  Yan C
Institution:School of Life Science and Engineering, Tsinghua University, Beijing, China.
Abstract:A pressurized gradient capillary electrochromatography (pCEC) instrument was developed to separate 18 amino acid derivatives. A reversed-phase C18 column (3 microm, 130 mm x 75 microm I.D.) and an acetate buffer (50 mmol/l NaAc, pH 6.4) with an ion-pair reagent (1% N,N-dimethylformamide) were used to separate derivatized amino acids from a standard solution (2 microg/ml), and the wavelength of the UV-Vis detector was 360 nm. The pressure on the capillary column was kept at approx. 70 Pa and 3 kV positive voltage was added on the outlet end of column. The effect of voltage on the eluting order of amino acids and the resolution of separation were studied, and it was found that when the voltage was higher than 3 kV, the adsorption of amino acids in the porous C18 column occurred. The effect of salt concentration, injection volume, and column length on the separation of amino acids was determined. The amino acid sample was separated by pCEC, and RSDs of the migration times of each amino acid were all less than 2.5%.