首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development of a capillary zone electrophoresis-electrospray ionisation tandem mass spectrometry method for the analysis of fluoroquinolone antibiotics
Authors:McCourt Josephine  Bordin Guy  Rodríguez Adela Rosa
Institution:

Institute for Reference Materials and Measurements, European Commission—Joint Research Centre (EC-JRC), Retieseweg, 2440, Geel, Belgium

Abstract:The applicability of a capillary zone electrophoresis–electrospray ionisation tandem mass spectrometric (CZE–ESI-MS–MS) method for the separation of nine fluoroquinolones was investigated. Method optimisation involved systematic trouble-shooting starting with the type and duration of capillary pre-washing and conditioning, the choice of both the CE run buffer, MS sheath liquid, CE run potential, ESI spray voltage, sheath gas flow-rate, MS capillary voltage and CE capillary and MS capillary temperatures. Another extremely important factor was found to be the degree to which the CE capillary protrudes into the ESI chamber as well as whether or not sheath gas and spray voltage are employed during the CE injection or not. The importance of the latter has, to our knowledge, not been addressed elsewhere. Nine fluoroquinolones have been separated and detected in a single run by this technique.
Keywords:Fluoroquinolones  Quinolones  Antibiotics
本文献已被 ScienceDirect PubMed 等数据库收录!