首页 | 本学科首页   官方微博 | 高级检索  
     检索      

甲苯中微量金属元素测定及特征分析
引用本文:李俊飞,李景喜,陈军辉,郑立,韩彬,王志广,李昉.甲苯中微量金属元素测定及特征分析[J].分析测试学报,2012,31(2):164-168.
作者姓名:李俊飞  李景喜  陈军辉  郑立  韩彬  王志广  李昉
作者单位:1. 国家海洋局第一海洋研究所海洋生态研究中心,山东青岛266061;青岛科技大学,化学与分子工程学院山东青岛266042
2. 国家海洋局第一海洋研究所海洋生态研究中心,山东青岛,266061
3. 青岛科技大学,化学与分子工程学院山东青岛,266042
基金项目:海洋公益性行业科研专项(201005034-3);国家海洋局青年基金资助项目(2011524)
摘    要:利用乙醇-水-稀硝酸乳化体系分散溶解甲苯样品,建立了电感耦合等离子体质谱(ICP-MS)同时测定甲苯中V、Cr、Mn、Fe、 Co、Ni、Cu、Zn、Ag、Cd及Pb等微量金属元素的新方法。采用有机进样系统,以Re多元素混标为内标元素,优化了仪器射频功率、载气流速、采样深度等仪器参数,以及乙醇-水比例和硝酸含量等样品前处理条件。结果表明,在最佳实验条件下,甲苯中多数微量金属元素的检出限均可达 ng/L级,各元素的线性关系良好(线性系数r=0.998 4~1.000);方法精密度较高,相对标准偏差(RSD)不大于8.0%;准确度和重现性较好,相对误差不高于10%。对不同厂家的甲苯样品进行分析,结果发现不同厂家样品的Zn、Mn与Co元素差异较大。该方法能够对不同厂家来源的甲苯进行区分,从而为实现海洋泄漏污染工业苯系物的鉴别提供一种新的辅助分析方法。

关 键 词:甲苯  微量金属元素    电感耦合等离子体质谱法

Determination of Trace Metal Elements in Toluene and Their Feature Analysis
LI Jun-fei , LI Jing-xi , CHEN Jun-hui , ZHENG Li , HAN Bin , WANG Zhi-guang , LI Fang.Determination of Trace Metal Elements in Toluene and Their Feature Analysis[J].Journal of Instrumental Analysis,2012,31(2):164-168.
Authors:LI Jun-fei  LI Jing-xi  CHEN Jun-hui  ZHENG Li  HAN Bin  WANG Zhi-guang  LI Fang
Institution:LI Jun-fei1,2,LI Jing-xi1*,CHEN Jun-hui1,ZHENG Li1,HAN Bin1,WANG Zhi-guang1,2,LI Fang2(1.Marine Ecology Research Center,First Institute of Oceanography of State Oceanic Administration, Qingdao 266061,China;2.College of Chemistry and Molecular Engineering,Qingdao University of Science & Technology,Qingdao 266042,China)
Abstract:An inductively coupled plasma mass spectrometric(ICP-MS) method was established for the determination of trace metal elements,including V,Cr,Mn,Fe,Co,Ni,Cu,Zn,Ag,Cd and Pb in toluene.The toluene were dissolved in ethanol-water-nitric acid system.The organic sampling system and the internal standard element Re were used in this experiment.The experimental conditions,such as RF power,carrier gas,sampling depth,ethanol-water ratio and nitric acid concentration were optimized.The results showed that,under the optimal conditions,the detection limits for the most trace metal elements could reach up to ng /L level.The linear correlation of the method was preferable with correlation coefficient of 0.998 4-1.000.The relative standard deviation(n=3) was less than 8.0% and the relative error was not more than 10%.The toluenes from different factories were analyzed.Results indicated that the concentrations of trace metal elements in the toluenes were obvious different.Therefore,a new identification method for benzenes spilled in ocean was provided by feature analysis of metals in the benzene compounds.
Keywords:toluene  trace metal elements  inductively coupled plasma mass spectrometry
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《分析测试学报》浏览原始摘要信息
点击此处可从《分析测试学报》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号