气相色谱法测定污水处理产生的氧 亚氮的分析方法研究
引用本文:郭刚,王亚宜,张兆祥,楚文海,潘绵立.气相色谱法测定污水处理产生的氧 亚氮的分析方法研究[J].分析测试学报,2011,30(9):1050-1054.
作者姓名:郭刚  王亚宜  张兆祥  楚文海  潘绵立
作者单位:1. 同济大学污染控制与资源化研究国家重点实验室,上海,200092
2. 上海医药工业研究院药理评价研究中心,上海,200040
基金项目:国家自然科学基金资助项目(51078283);全国优秀博士学位论文作者专项基金资助(200756);国家教育部新世纪优秀人才支持计划资助项目(NCET-08-0404);上海市科技人才计划资助项目(09QA1406100)
摘    要:建立了气相色谱法(GC/ECD)检测污水处理过程中产生的温室气体氧化亚氮(N2O)的分析方法.确定了最佳实验条件:进样口温度、柱温和ECD检测器温度分别为105、100、300℃,以95% Ar +5% CH4为载气,柱流速和尾吹气流速分别为25、50 mL/min.N2O的质量浓度在0.255~ 100 mg/L范围...

关 键 词:氧化亚氮  气相色谱  电子捕获器  温度  载气  灵敏度

A Gas Chromatographic Method for Determination of Nitrous Oxide in Wastewater Treatment
GUO Gang,WANG Ya-yi,ZHANG Zhao-xiang,CHU Wen-hai,PAN Mian-li.A Gas Chromatographic Method for Determination of Nitrous Oxide in Wastewater Treatment[J].Journal of Instrumental Analysis,2011,30(9):1050-1054.
Authors:GUO Gang  WANG Ya-yi  ZHANG Zhao-xiang  CHU Wen-hai  PAN Mian-li
Institution:GUO Gang1,WANG Ya-yi1,ZHANG Zhao-xiang1,CHU Wen-hai1,PAN Mian-li2(1.State Key Laboratory of Pollution Control and Resources Reuse,Tongji University,Shanghai 200092,China,2.Center for Pharmacological Evaluation,Shanghai Institute of Pharmaceutical Industry,Shanghai 200040,China)
Abstract:A gas chromatographic(GC) method was established for the analysis of a potential greenhouse gas,nitrous oxide(N2O) in wastewater treatment.The optimum conditions of chromatography were determined by a series of parallel tests.The working temperatures of the injector,column and electron capture detector(ECD) were 105,100 and 300 ℃,respectively;95% Ar +5% CH4 was selected as the carrier gas;the column flow and the makeup gas flow were 25 mL/min and 50 mL/min,respectively.Under the optimal conditions,the calibration curve was linear in the range of 0.255-100 mg/L N2O,with correlation coefficient higher than 0.999 and the mean recovery larger than 90%.When detecting 100 mg/L N2O(standard samples) using direct injection and headspace methods,the relative standard deviation(RSD) values were 1.5% and 0.32%,respectively,and the recoveries were both over 90%.The method was successfully applied in the determination of N2O in wastewater treatment processes and the detected values are consistent with the results reported in literatures.This method is simple,sensitive and repeatable.
Keywords:nitrous oxide  gas chromatography  electron capture detector  temperature  carrier gas  sensitivity  