首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Novel ibuprofen potentiometric membrane sensors based on tetraphenylporphyrinato indium(III).
Authors:Saad S M Hassan  Wagiha H Mahmoud  Mohamed A F Elmosallamy  Mahra H Almarzooqi
Institution:Department of Chemistry, Faculty of Science, Ain Shams University, Cairo, Egypt. saadsmhassan@yahoo.com
Abstract:Two novel potentiometric membrane sensors responsive to the ibuprofen drug have been developed. These incorporate poly(vinyl chloride) and polyurethane matrix membranes containing 5,10,15,20-tetraphenylporphrinato (TPP) indium(II) ionophore plasticized with dibutylsebacate. The sensors show a near-Nernstian response with anionic slopes of -53 and -55 mV decade(-1), over the concentration range of 4.2 x 10(-6)-1.0 x 10(-2) and 3.3 x 10(-6)-1.0 x 10(-2) M ibuprofen within pH ranges of 4-9 and 5-9 for PVC and PU matrix membranes, respectively. A sensor based on a polyurethane membrane displays a lower detection limit and a wider linear working range, and a sensor based on a PVC membrane exhibits a better overall selectivity, especially in the presence of lipophilic organic anions. Both sensors are used for the quantification and quality-control assessment of ibuprofen in pharmaceutical preparations. The average recoveries are 99.1+/-0.3% and 99.3+/-0.3% for TPP In(III)-PVC and TPP In(III)-PU based membrane sensors, respectively. High selectivities towards ibuprofen in the presence of many anions, drug excipients and diluents are offered by both sensors, which exhibit a non-Hofmeister selectivity pattern.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号