首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Silver(I)-selective coated-wire electrode based on an octahydroxycalix[4]arene derivative.
Authors:Mohammad Mazloum Ardakani  Ali Asghar Ensafi  Mohammad Salavati Niasari  Hossain Mirhoseini
Institution:Department of Chemistry, Kashan University, Kashan, Iran.
Abstract:The performance of octahydroxycalix4]arene derivative used as a neutral carrier for silver polymeric membrane electrode was studied. The sensor gave a good Nernstian response of 58 +/- 1 mV per decade for silver ion in the activity range 3.3 x 10(-6) to 3.3 x 10(-2) M Ag+. The limit of detection reached 2.1 x 10(-6) M Ag+ and exhibited high selectivity for silver ion against the alkali, alkaline earth and transition metal ions. The sensor can be used in wide pH range from 1.5 to 6.5. The response time of the sensor is less than 20 s. The potentiometric sensor was used as the indicator electrode in the titration of Ag+ ions by sodium chloride solution.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号