Lead(II)-selective electrode based on phenyl disulfide.
Authors:Abdolkarim Abbaspour  Behnam Khajeh
Institution:Chemistry Department, Shiraz University, Iran.
Abstract:A PVC membrane electrode for lead ions based on phenyl disulfide as the membrane carrier was developed. The electrode exhibits a good Nernstian slope of 29.3 +/- 0.7 mV/decade and a linear range of 2.0 x 10(-6)-1.0 x 10(-2) M for Pb(NO3)2. The limit of detection is 1.2 x 10(-6) M. It has a response time of 45 s and can be used for at least fifty days without any divergence in potential. The proposed membrane sensor revealed high selectivity for Pb2+ over a wide variety of other metal ions and could be used in the pH range of 3.5-6.3. The electrode was used as an indicator electrode in potentiometric titration of lead ions.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏