首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Poly(dimethylsiloxane) microchip capillary electrophoresis with electrochemical detection for rapid measurement of acetaminophen and its hydrolysate
Authors:F Y He  A L Liu  X H Xia
Institution:(1) The State Key Laboratory of Coordination Chemistry, Institute of Analytical Science, Department of Chemistry, Nanjing University, Nanjing, 210093, China
Abstract:Poly(dimethylsiloxane) microchip capillary electrophoresis with amperometric detection has been used for rapid separation and determination of acetaminophen and its hydrolysate, i.e. p-aminophenol. A Pt ultramicroelectrode with a diameter of 10 mgrm positioned at the outlet of the separation channel was used as a working electrode for amperometric detection. Factors influencing separation and detection were investigated and optimized. Results show that acetaminophen and p-aminophenol can be well separated within 35 s with RSD<1% for migration time and <7% for detection current for both analytes. Detection limits for both analytes are estimated to be 5.0 mgrmol L–1 (approximately 0.1 fmol) at S/N=3. This method has been successfully applied to the detection of traces of p-aminophenol in paracetamol tablets.
Keywords:Poly(dimethylsiloxane)  Microchip capillary electrophoresis  Electrochemical detection  Acetaminophen  p-Aminophenol  Ultramicroelectrode  Paracetamol tablet
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏