首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Application of transmission near-infrared spectroscopy to uniformity of content testing of intact steroid tablets
Authors:Broad N W  Jee R D  Moffat A C  Smith M R
Institution:Centre for Pharmaceutical Analysis, The School of Pharmacy, University of London, UK.
Abstract:Transmission near-infrared (NIR) spectroscopy was used for the rapid and non-destructive determination of the content of a hormone steroid in single intact tablets. Tablets produced for clinical trial purposes containing 5, 10, 15, 20 and 30 mg (2.94, 5.88, 8.82, 11.76 and 17.64% m/m, respectively) were used to develop calibration models without the need to specially prepare any out of specification tablets. Reference values for the individual tablets used in the NIR calibration models and test set were measured by reversed-phase high performance liquid chromatography (HPLC). Partial least squares regression using standard normal variate transformed second-derivative spectra over the range 800 to 1040 nm gave the optimum calibration model with a standard error of calibration of 0.52 mg per tablet. Measurements of an independent test set gave comparable results (standard error of prediction 0.31 mg per tablet). Measurement errors for a single tablet (RSD < 2.5% for a given active level) were sufficiently small to allow the procedure to be applied to pharmacopoeial uniformity of content testing of batches of these tablets and permitted the non-destructive testing of 30 tablets in under 20 min as compared to 6 h by HPLC.
Keywords:
本文献已被 PubMed 等数据库收录!