Abstract: | Near infrared spectroscopy in combination with appropriate chemometric methods is an effective technique for quantitative analysis of parameters of interest for the pharmaceutical industry. In this study, the artificial neural network (ANN) was applied to monitor critical parameters (compression force, tablet hardness, mean particle size, and active pharmaceutical ingredient concentration of tablets) in the process of naproxen pharmaceutical preparation. The performance of ANN was compared to linear methods (partial least squares regression (PLS) and synergy interval partial squares (siPLS)). The ANN models for compression force, tablet hardness, mean particle size, and active pharmaceutical ingredient concentration of tablets yielded the low root mean square error of prediction (RMSEP) values of 0.936 KN, 0.302 kg, 4.49 mg, and 2.14 µm, respectively. The predictive ability of the PLS model was improved by siPLS with selection of spectral regions and the best performance among all calibration methods was showed by the nonlinear method (ANN). Effective models were built by using these approaches using near infrared spectroscopy. |