首页 | 本学科首页   官方微博 | 高级检索  
     检索      

流动注射双安培法测定多巴胺
引用本文:李利军,程昊,陈其锋,黄文艺,孔红星,吴健玲.流动注射双安培法测定多巴胺[J].分析化学,2006,34(8):1129-1132.
作者姓名:李利军  程昊  陈其锋  黄文艺  孔红星  吴健玲
作者单位:1. 广西工学院生物与化学工程系,柳州,545006
2. 广西大学化学与化学工程学院,南宁,530004
基金项目:广西青年科学基金(桂科青0135003),广西高校百名中青年学科带头人资助计划(桂教人[2002]467)资助项目
摘    要:通过偶合多巴胺在铂电极上的氧化和高锰酸钾在铂电极上的还原,建立了一个不施加电压的条件下的流动注射双安培法直接测定多巴胺的新方法。以0.05 mol/L硫酸为载液,多巴胺的氧化峰电流与其浓度在0.8~160 mg/L范围内呈线性关系,线性回归方程为i(nA)=652.9C-239.2(r=0.9998,n=10),检出限为0.2 mg/L;RSD为2.86%(N=80 mg/L,n=14);进样频率为80次/h。本方法具有很高的选择性和灵敏度,样品处理方法简单快速,适于连续自动测定。用于实际样品的测定,结果满意。

关 键 词:多巴胺  流动注射  不可逆双安培  铂电极
收稿时间:11 15 2005 12:00AM
修稿时间:2005-11-152006-03-09

Determination of Dopamine by Flow Injection-Irreversible Biamperometric
Li Lijun,Cheng Hao,Chen Qifeng,Huang Wenyi,Kong Hongxing,Wu Jianling.Determination of Dopamine by Flow Injection-Irreversible Biamperometric[J].Chinese Journal of Analytical Chemistry,2006,34(8):1129-1132.
Authors:Li Lijun  Cheng Hao  Chen Qifeng  Huang Wenyi  Kong Hongxing  Wu Jianling
Institution:1. Department of Biological and Chemical Engineering, Guangxi University of Technology, Liuzhou 545006 ; 2. College of Chemistry and Chemical Engineering, Guangxi University, Nanning 530004
Abstract:A simple flow-injection irreversible biamperometric method for the determination of dopamine hydrochloride injection has been described.The method is based on the electrocatalytic oxidation of dopamine hydrochloride at one(platinum) electrode,which is coupled with the reduction of permanganate at another(platinum) electrode by external circuitry,composing such a biamperometric system as a pseudo-reversible one.The system can run without the application of an external potential.The oxidation current of dopamine hydrochloride shows linear relationship from 0.8 mg/L to 160 mg/L with a detection limit of 0.2 mg/L.The(relative) standard deviation(RSD) of the method is 2.9% for 14 reduplicate determination of 80 mg/L dopamine hydrochloride.The method offers the advantages of simplicity,high efficiency(80 samples/h),and high selectiveity,and is suitable for the determination of dopamine hydrochloride in injection.
Keywords:Dopamine  flow injection  irreversible biamperometry  platinum electrode
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号