Neutron dosimetry device using PADC nuclear track detectors
Authors:Gisela Saint Martin  F López  O A Bernaola
Institution:1.Lab. Tandar. Radiobiología. Comisión Nacional de Energía Atómica,Buenos Aires,Argentina;2.Instituto J. Sábato, UNSAM,Buenos Aires,Argentina
Abstract:A device consisting of poly allyl diglycol carbonate (PADC) foils with a polyethylene and polyvinyl chloride (PVC) radiator was calibrated in order to perform the dosimetry of an 241Am-Be source covering a dose range from 0 to 3.1 mSv. Working conditions were established for a good visualization and counting of tracks: chemical attack was made in two steps with different solutions and etching times. A good fit was obtained for the dose as a function of the number of tracks per unit surface area. The calculated decision threshold was 0.12 mSv.
Keywords:
本文献已被 SpringerLink 等数据库收录!