Stripping voltammetric detection of mercury(II) based on a surface ion imprinting strategy in electropolymerized microporous poly(2-mercaptobenzothiazole) films modified glassy carbon electrode
Authors:Fu Xu-Cheng  Chen Xing  Guo Zheng  Xie Cheng-Gen  Kong Ling-Tao  Liu Jin-Huai  Huang Xing-Jiu
Institution:aResearch Center for Biomimetic Functional Materials and Sensing Devices, Institute of Intelligent Machines, Chinese Academy of Sciences, Hefei, Anhui, 230031, PR China;bAnhui Provincial Laboratory of Biomimetic Sensor and Detecting Technology, West Anhui University, Lu’an, Anhui, 237015, PR China
Abstract:This work reports a surface ion imprinting strategy in electropolymerized microporous poly(2-mercaptobenzothiazole) (MPMBT) films at the surface of glassy carbon electrode (GCE) for the electrochemical detection of Hg(II). The Hg(II)-imprinted MPMBT/GCE exhibits larger binding to functionalized capacity, faster binding kinetics and higher selectivity to template Hg(II) due to their high ratio of surface-imprinted sites, larger surface-to-volume ratios, the complete removal of Hg(II) templates and larger affinity to Hg(II). The square wave anodic stripping voltammetry (SW ASV) response of the Hg(II)-imprinted MPMBT/GCE to Hg(II) is ca. 3.0 and 5.9 times larger than that at the direct imprinted poly(2-mercaptobenzothiazole) modified GCE and non-imprinted MPMBT/GCE sensor, respectively; and the detection limit for Hg(II) is 0.1 nM (which is well below the guideline value given by the World Health Organization). Excellent wide linear range (1.0–160.0 nM) and good repeatability (relative standard deviation of 2.5%) were obtained for Hg(II). The interference experiments showed that mercury signal was not interfered in the presence of Pb(II), Cd(II), Zn(II), Cu(II) and Ag(I), respectively. These values, particularly the high sensitivity and excellent selectivity compared favorably with previously reported methods in the area of electrochemical Hg(II) detection, demonstrate the feasibility of using the prepared Hg(II)-imprinted MPMBT/GCE for efficient determination of Hg(II) in aqueous environmental samples.
Keywords:Surface ion imprinting  Square wave anodic stripping voltammetry  Electropolymerization  Microporous film  Mercury(II) detection
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号