Calibration methods for microdialysis sampling in vivo: muscle and adipose tissue
Authors:Yuan Song  Craig E Lunte  
Institution:

Department of Chemistry, University of Kansas, Lawrence, KS 66045, USA

Abstract:Calibration methods for microdialysis sampling were studied in the muscle and adipose tissue of rats. Both the delivery method and the no-net-flux method were used to determine the extraction efficiency (EE) of acetaminophen and caffeine in both tissues. There was no concentration dependence of the EE either in vitro or in vivo for either acetaminophen or caffeine. The EEs determined by the delivery and no-net-flux methods were not different. However, the EEs of both caffeine and acetaminophen determined in vitro were significantly higher than those determined in the muscle and adipose. This indicates that mass transfer in the tissue is the rate-determining factor for the EE in vivo. The relative difference between the EE in vitro and the EE in the muscle was smaller than the difference between the EE in vitro and the EE in the adipose. In addition, the EE in the muscle decreased more than the EE in the adipose after the animal was euthanized. This indicated that exchange between the extracellular fluid and plasma is the rate-determining step in mass transport relative to microdialysis sampling. This has a more significant effect on the EE in the muscle than the EE in the adipose. Both the delivery and no-net-flux methods can be used to calibrate microdialysis probes in the muscle and adipose.
Keywords:Microdialysis  Calibration  In vivo analysis  Anatomy