首页 | 本学科首页   官方微博 | 高级检索  
     检索      

SMALL-BORE COLUMNS OF 1 mm I. D. FOR HPLC
作者姓名:卢佩章  张玉奎  包绵生  周桂敏  杨春和  林从敬
作者单位:Dalian Institute of Chemical Physics,Academia Sinica,Dalian Institute of Chemical Physics,Academia Sinica,Dalian Institute of Chemical Physics,Academia Sinica,Dalian Institute of Chemical Physics,Academia Sinica,Dalian Institute of Chemical Physics,Academia Sinica,Dalian Institute of Chemical Physics,Academia Sinica
摘    要:The performance of small-bore columns of 1 mm I. D. for HPLC is described. The effect of mobile phase velocity and connecting effect on the values of H, a′, b′ and β are investigated. By operating at mobile phase velocity of 0.8 mm/s, the column of 200×1 mm I. D. packed with YWG-5μm (silica gel) can give the efficiency of about 11,000 plates which meet the needs of ordinary analysis. Oompared with column of 150×2 mm I. D., the solvent consumption is reduced to 1/3. The value of n~∞/t_r~0 about 700 T. P./s can be obtained. But the value of β is rather large. It needs further to be improved. Coupling 5 columns of 200×1 mm I. D., the efficiency of about 50,000 plates can be obtained. For a column of 1000×1 mm I. D. packed directly, the efficiency of 70,000 plates was achieved. Some applications of 1 mm I. D. column are given. We propose to use the column of 200×1mm I. D. in the routine analysis.