首页 | 本学科首页   官方微博 | 高级检索  
     检索      


An investigation of calibration materials for the measurement of metals in ambient particulate matter on filters by LA-ICP-MS
Authors:Richard JC Brown  Kim E Jarvis  Benoit A Disch  Sharon L Goddard
Institution:1. Analytical Science Team, National Physical Laboratory , Teddington , Middlesex, TW11 0LW , UK richard.brown@npl.co.uk;3. Centre for Environmental Policy, Imperial College , Manor House, Silwood Park, Ascot , SL5 7PY , UK;4. Analytical Science Team, National Physical Laboratory , Teddington , Middlesex, TW11 0LW , UK
Abstract:Three different types of simple and low-cost calibration material for the measurement of the metals content of ambient particulate matter (PM) on filters using laser ablation-inductively coupled plasma-mass spectrometry (LA-ICP-MS) have been compared: cellulose ester filters spiked with multi-element calibration solutions, pellets of compressed ambient particulate matter certified reference material (CRM), and powdered ambient particulate matter CRM adhered to a surface. Elements determined were As, Cd, Cr, Cu, Fe, Mn, Ni, Pb, V and Zn, each at approximate levels of 1000?ng per filter. Blank filters spiked with multi-elemental standards were found to be significantly more reproducible and repeatable than materials based on powdered reference materials. However, a comparison of these spiked filters with real samples of ambient PM showed that the analytical sensitivities obtained per mass of analyte were significantly different. It is concluded that the spiked filters could act as very effective quality control standards correcting, to within 1%, drifts in LA-ICP-MS measurements of up to 10%, or as indirect calibration materials supported by additional measurements using traditional wet chemical techniques.
Keywords:ambient air  particulate matter  metals  calibration  LA-ICP-MS
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号