A calibration procedure for millimeter-scale stereomicroscopic particle image velocimetry
Authors:John Giardino  Jean Hertzberg  Elizabeth Bradley
Institution:(1) UCB 427, University of Colorado, Boulder, CO 80309-0427, USA;(2) Department of Mechanical Engineering, University of Colorado, Boulder, CO, USA;(3) Department of Computer Science, University of Colorado, Boulder, CO, USA
Abstract:A stereomicroscopic particle image velocimetry (SμPIV) system has been developed for millimeter scale flows. The SμPIV system is based on an off-the-shelf stereomicroscope, with magnification between 0.69× and 30×, and a field of view between 7.5 × 6 mm and 250 × 200 μm. Custom calibration targets were devised using printed circuit board technology, and applied at a magnification factor of 1.74, with a field of view of 4.75 × 3.8 mm. Measurement errors were assessed by moving a test block with fixed particles. Total system uncertainty in test block displacement transverse to the optical axis was 0.5% of the field of view, and 3% of the depth of field for motion along the optical axis. Approximately 20% of this uncertainty was due to the calibration target quality and test block procedures.