Possibilities and limits of sound power measurements with a real-time intensity analyzer |
| |
Authors: | A Stirnemann U Bolleter EJ Rathe |
| |
Institution: | Research Laboratory for Vibrations and Acoustics, Sulzer Brothers Limited, Winterthur, Switzerland;Swiss Federal Institute of Technology, Zurich, Switzerland |
| |
Abstract: | The sound power of a number of test objects was determined from spatially averaged intensity measurements. The results show that the influence of room acoustics is insignificant even for rooms of widely different room constants, if the measuring surfaces are exactly defined and if a good space-averaging technique is used. The intensity integrated over a closed surface defining a source-free space compared to the sound pressure integrated over the same surface gives a measure of the capability of a specific intensity measuring system to suppress external noise. For the test arrangements measured with broad band noise, this suppression was found to be 14–18 dB(A). A similar value of 15 dB was found from sound power measurements on a source with high external sound and an analysis of the results in one-third octave bands. From these measurements an analytical function was derived which describes the average error of the spatially averaged intensity as a function of the difference between the external sound level and the source sound level. For practical measurement situations a further analytical function was derived which gives this intensity error as a function of the difference between the measured (spatially averaged) pressure and intensity levels. Thus it is possible to estimate the error of intensity measurements directly from measured intensity and pressure data. |
| |