LS301快中子探测器的效率刻度
引用本文:张文慧,黎光武,张刚,陈宇,闫果红.LS301快中子探测器的效率刻度[J].原子核物理评论,2011,28(2):195-199.
作者姓名:张文慧  黎光武  张刚  陈宇  闫果红
作者单位:1 中国原子能科学研究院, 北京102413;2 中国科学院近代物理研究所, 甘肃 兰州730000;3 中国科学院高能物理研究所, 北京100049;4 山西师范大学现代物理研究所, 山西 临汾041004
基金项目:国家自然科学基金资助项目
摘    要:主要介绍了利用252Cf标准裂变中子能谱刻度快脉冲LS301型液闪中子探测器的探测效率的实验方法和结果, 简要介绍了实验数据的处理过程, 得到了阈值分别为0.5, 0.7, 1.0和1.6 MeV, 以及中子能量在10 MeV以下的探测器效率, 并对测量结果进行了误差分析。同时为了检验实验结果的准确性, 将实验结果与理论计算结果进行比较, 两者在不确定度范围内符合得很好。Neutron detection efficiency of LS301 fast neutron detector was calibrated by measuring the neutron energy spectrum of 252Cf source, which has a standard fission neutron spectrum. A low mass, fast ionization chamber is used as the fission fragments detector in the time of flight(TOF) spectrometer and afforded the start signal of neutron flight. The stop signal was offered by the anode of LS301. A measured TOF spectrum was turned to the neutron energy spectrum which will be compared with the standard one. Consequently, the fast neutron detection efficiency of LS301 was obtained. Calibration procedures of experimental and data processing was given. Relative detection efficiencies were obtained separately for threshold settings of 0.5, 0.7, 1.0 and 1.6 MeV for neutron energies under 10 MeV. Experimental results accorded with theoretical efficiency curves which were calculated with the Monte Carlo code NEFF50.

关 键 词:252Cf电离室    LS301快中子探测器    中子探测效率    飞行时间
收稿时间:1900-01-01

Calibration of LS301 Fast-neutron Detector Efficiency
ZHANG Wen-hui,LI Guang-wu,ZHANG Gang,CHEN Yu,YAN Guo-hong.Calibration of LS301 Fast-neutron Detector Efficiency[J].Nuclear Physics Review,2011,28(2):195-199.
Authors:ZHANG Wen-hui  LI Guang-wu  ZHANG Gang  CHEN Yu  YAN Guo-hong
Institution:1 China Institute of Atomic Energy, Beijing 102413, China;2 Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China;3 Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China;4 Institute of Modern Physics, Shanxi Normal University, Linfen 041004, Shanxi, China
Abstract:Neutron detection efficiency of LS301 fast neutron detector was calibrated by measuring the neutron energy spectrum of 252Cf source, which has a standard fission neutron spectrum. A low mass, fast ionization chamber is used as the fission fragments detector in the time of flight(TOF) spectrometer and afforded the start signal of neutron flight. The stop signal was offered by the anode of LS301. A measured TOF spectrum was turned to the neutron energy spectrum which will be compared with the standard one. Consequently, the fast neutron detection efficiency of LS301 was obtained. Calibration procedures of experimental and data processing was given. Relative detection efficiencies were obtained separately for threshold settings of 0.5, 0.7, 1.0 and 1.6 MeV for neutron energies under 10 MeV. Experimental results accorded with theoretical efficiency curves which were calculated with the Monte Carlo code NEFF50.
Keywords: