首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Variable square pulse vs conventional PFN pumping of Er:YAG laser
Authors:K Neme?  M Luka?  J Mo?ina
Institution:1. Fotona d.d., Stegne 7, 1210 Ljubljana, Slovenia;2. Institute Josef Stefan, Light and Matter Department, Jamova 39, 1000 Ljubljana, Slovenia;3. University of Ljubljana, Faculty of Mechanical Engineering, A?ker?eva 6, 1000 Ljubljana, Slovenia
Abstract:The influence of the flashlamp pump current pulse shape on Er:YAG laser efficiency and laser rod thermal focusing was studied theoretically and experimentally. Two pulse shapes, PFN (Pulse Forming Network) and VSP (Variable Square Pulse), were considered. Theoretical modeling and experimental measurements show that the pump pulse shape itself does not have a significant influence on the Er:YAG laser efficiency or thermal focusing. Instead, the major parameter influencing Er:YAG laser efficiency and thermal focusing was found to be the overall pulse duration. For PFN pulses, rise and fall times directly define the overall pulse duration, and therefore do have influence on thermal focusing. By contrast, VSP pulse duration is defined by the externally controlled on-time of the switching transistor. For square shaped pulses, short rise and fall times do not have a direct beneficial influence on thermal lensing.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号