首页 | 本学科首页   官方微博 | 高级检索  
     检索      


In situ lift-out using a FIB-SEM system
Authors:Langford R M  Clinton C
Institution:SFI Trinity Nanoscience Laboratory, Department of Physics, Trinity College Dublin, Dublin 2, Ireland. richard.langford@tcd.ie
Abstract:An in situ 'lift-out' technique for preparing transmission electron microscopy specimens, which is performed using secondary electron imaging within the chamber of a focused ion beam-SEM system is presented. The main advantage of this in situ approach, relative to the ex situ lift-out technique, is that secondary electron imaging enables higher magnifications to be used than is possible with optical microscopy. This makes the lift-out procedure more controllable and thus increases the overall success rate especially for inexperienced users. The technique is also compared with another in situ lift-out approach, the 'wedge' technique.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏