经验模态分解法在近红外无创血红蛋白检测中的应用研究
引用本文:樊奕辰,卢启鹏,丁海泉,高洪智,陈星旦.经验模态分解法在近红外无创血红蛋白检测中的应用研究[J].光谱学与光谱分析,2013,33(2):349-353.
作者姓名:樊奕辰  卢启鹏  丁海泉  高洪智  陈星旦
作者单位:1. 中国科学院长春光学精密机械与物理研究所,应用光学国家重点实验室,吉林 长春 130033
2. 中国科学院研究生院,北京 100049
基金项目:国家自然科学基金项目(60938002);应用光学国家重点实验室开放基金项目(Y1Q03FQ113);中国科学院知识创新工程领域前沿项目(Y00232Q100)资助
摘    要:为提高近红外血红蛋白预测模型的稳健性,分别应用Savitzky-Golay平滑、移动窗口平滑以及经验模态分解(EMD)方法对原始光谱进行去噪处理,以提高数据信噪比。采集了81例临床志愿者的手指指端血流容积脉搏波光谱数据,同时获取相应的血红蛋白浓度值临床化验结果。剔除异常样品,确定78例样品为研究对象,建立反向传播神经网络(BP-ANN)定量分析模型并预测。结果表明,经EMD处理后的模型预测效果最优,预测相关系数由0.74提高至0.87,误差均方根由12.85 g·L-1减小至8.08 g·L-1。实验证明应用EMD方法能够获得高信噪比的容积脉搏信号,提高血红蛋白浓度预测模型的准确性,有利于推动近红外无创血红蛋白检测技术的进一步发展。

关 键 词:近红外光谱  血红蛋白浓度  经验模态分解  无创检测    
收稿时间:2012-06-01

Study on the Application of Empirical Mode Decomposition to Noninvasive Hemoglobin Measurement by NIRS
FAN Yi-chen,LU Qi-peng,DING Hai-quan,GAO Hong-zhi,CHEN Xing-dan.Study on the Application of Empirical Mode Decomposition to Noninvasive Hemoglobin Measurement by NIRS[J].Spectroscopy and Spectral Analysis,2013,33(2):349-353.
Authors:FAN Yi-chen  LU Qi-peng  DING Hai-quan  GAO Hong-zhi  CHEN Xing-dan
Institution:1. State Key Lab of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China2. Graduate University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:To increase the signal-to-noise ratio (SNR) of human near infrared (NIR) spectra, so as to improve the stability and precision of calibration model, the empirical mode decomposition (EMD) method was applied. Eighty-one fingertip absorption curves were collected, with the corresponding clinical examination results obtained immediately. By means of outliers detection and removal, finally 78 samples were determined as the research objects. A three-layer back-propagation artificial neutron network (BP-ANN) model was established and worked for prediction. The results turned out that, through EMD method, the prediction correlation coefficient increased greatly from 0.74 to 0.87. RMSEP was reduced from 12.85 to 8.08 g·L-1. Other indexes were also obviously improved. The overall results sufficiently demonstrate that it is feasible to use EMD method for high SNR pulse wave signals, thus improving the performance of noninvasive hemoglobin calibration models. The application of EMD method can help promote the development of noninvasive hemoglobin monitoring technology.
Keywords:Near infrared spectroscopy  Hemoglobin concentration  Empirical mode decomposition  Noninvasive measurement  