首页 | 本学科首页   官方微博 | 高级检索  
     检索      

流动注射-可见光与原子吸收联机测定血清中磷及钙
引用本文:张宏绪,马宾胜,李秀荣.流动注射-可见光与原子吸收联机测定血清中磷及钙[J].光谱学与光谱分析,1994(1).
作者姓名:张宏绪  马宾胜  李秀荣
作者单位:哈尔滨医科大学公共卫生学院
摘    要:采用流动注射一次进样串联测定血清中磷、钙,即以串联的可见分光光度计与原子吸收分光光度计先后连续测定磷和钙。试验了流动注射系统的实验条件。结果磷、钙回收率分别为99.7%、102.0%,变异系数各为2.7%、2.5%,即与手动法结果相同或更优

关 键 词:流动注射,磷和钙,串联测定

DETERMINATION OF PHOSPHORUS AND CALCIUM IN SERUM BY FLOW INJECTION WITH SERIES DETECTORS OF SPECTROPHOTOMETER AND ATOMIC ABSORPTION
ZHANG Hongxu, MA Binsheng and LI Xiurong.DETERMINATION OF PHOSPHORUS AND CALCIUM IN SERUM BY FLOW INJECTION WITH SERIES DETECTORS OF SPECTROPHOTOMETER AND ATOMIC ABSORPTION[J].Spectroscopy and Spectral Analysis,1994(1).
Authors:ZHANG Hongxu  MA Binsheng and LI Xiurong
Institution:ZHANG Hongxu; MA Binsheng and LI Xiurong
Abstract:Phosphorus and Calcium in serum were measured succesively by the method of flow injection with series detectors of spectrophotometer and Atomic Absorption. Each sample was put in at one time. During the measurement, the conditions for the Flow-injection system were chosen. The results show that the recoveries of phosphorus and Calcium are 99. 7% and 102. 0% respectively, and Coefficients of variability are 2. 7 % and 2. 5 % respectively. It is indicated that this method is equal to or better than that of manual work.
Keywords:Flow injection  Phosphorus and Calcium  Series detector
本文献已被 CNKI 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号