悬浮液进样端视电感耦合等离子体发射光谱法直接测定二氧化钛中铌
引用本文:汪正,陈天裕,陶光仪,杨芃原.悬浮液进样端视电感耦合等离子体发射光谱法直接测定二氧化钛中铌[J].光谱学与光谱分析,2005,25(4):556-559.
作者姓名:汪正  陈天裕  陶光仪  杨芃原
作者单位:1. 中国科学院上海硅酸盐研究所,上海 200050
2. 复旦大学化学系,上海 200433
基金项目:上海科学技术发展基金 (0 1 59nm 0 75)资助项目
摘    要:用聚丙烯酸胺(NH4PAA)为分散剂制备二氧化钛悬浮液,并以悬浮液进样端视电感耦合等离子体发射光谱法直接测定了二氧化钛中痕量杂质铌。研究了分散剂及分散剂的量和pH对二氧化钛悬浮液稳定性的影响,通过优化条件制备得到稳定和均匀的悬浮液。研究了端视电感耦合等离子体发射光谱分析悬浮液进样的性能。本方法的检出限为3.0 μg·L-1, 相对标准偏差为3.1%(n=3, c=0.3 mg·L-1)。

关 键 词:悬浮液  端视  电感耦合等离子体发射光谱  二氧化钛    
文章编号:1000-0593(2005)04-0556-04
收稿时间:2003-09-06
修稿时间:2003年9月6日

Direct Determination of Nb in Titanium Oxide Using Slurry Sample Introduction by Axial Viewing Inductively Coupled Plasma Optical Emission Spectrometry
WANG Zheng,CHEN Tian-Yu,TAO Guang-yi,YANG Peng-yuan.Direct Determination of Nb in Titanium Oxide Using Slurry Sample Introduction by Axial Viewing Inductively Coupled Plasma Optical Emission Spectrometry[J].Spectroscopy and Spectral Analysis,2005,25(4):556-559.
Authors:WANG Zheng  CHEN Tian-Yu  TAO Guang-yi  YANG Peng-yuan
Institution:1. Shanghai Institute of Ceramics, Chinese Academy of Sciences, Shanghai 200050, China2. Chemistry Department, Fudan University, Shanghai 200433, China
Abstract:By using dispersant polyacrylate amine (NH4PAA) to disperse TiO2, effects of different dispersants, pH value and the amount of dispersant on TiO2 slurry were investigated. The pH value of the medium and the dispersant amount were optimized, and a stable and homogeneous suspension was prepared. Nb in TiO2 was determined by an axial viewing ICP-OES spectrometer. At the same time, the performance of the axial viewing ICP-OES using solid powder analysis was discussed. Under the optimum experimental conditions, the detection limit of the present method is 3.0 microg x L(-1) and the RSD is 3.1% (n = 3, c = 0.3 mg x L(-1)).
Keywords:Slurry  Axial  ICP-OES  Ti_2  Nb