样品装样、测试条件等因素对近红外检测结果的影响与分析误差源比较研究
引用本文:李军会,秦西云,张文娟,蔡贵民,杨宇虹,赵龙莲,常志强,赵丽丽,张录达.样品装样、测试条件等因素对近红外检测结果的影响与分析误差源比较研究[J].光谱学与光谱分析,2007,27(9):1751-1753.
作者姓名:李军会  秦西云  张文娟  蔡贵民  杨宇虹  赵龙莲  常志强  赵丽丽  张录达
作者单位:1. 中国农业大学信息与电气工程学院,北京 100094
2. 云南省烟草科学研究所,云南 玉溪 653100
3. 中国农业大学理学院,北京 100094
基金项目:云南省科技厅科技攻关重大项目 , 国家科技攻关项目 , 国家高技术研究发展计划(863计划)
摘    要:以云南优质烤烟为实验材料,在国产光栅漫反射型近红外仪器上,研究了样品装样、测试条件等因素对近红外检测结果的影响。并实现了通过建立全局模型应用自校正方法来降低分析误差。结果表明,在不考虑样品粒度、测试条件等因素的影响情况下,该仪器自身信噪比等性能指标引起的测试重复性误差、重装样误差、样品松紧度误差来源各占50%,30%,20%左右,通过压样措施可降低由样品松紧度产生的误差。当季节、温度等测试条件发生较大变化时,引起的测试准确性误差远大于所有测试重复性误差的总和。文章旨在从误差来源理论上解析各影响因素,为改进仪器提供基础理论数据;从理论上解析测试条件因素影响的重要性,并为近红外技术研究解决该问题给出了一种新思路。

关 键 词:近红外  烤烟  装样  测试条件  误差来源  
文章编号:1000-0593(2007)09-1751-03
收稿时间:2006-05-19
修稿时间:2006-05-19

Influence of Sample Loading and Test Conditions on NIR Veracity and Study of Analysis Error Source
LI Jun-hui,QIN Xi-yun,ZHANG Wen-juan,CAI Gui-min,YANG Yu-hong,ZHAO Long-lian,CHANG Zhi-qiang,ZHAO Li-li,ZHANG Lu-da.Influence of Sample Loading and Test Conditions on NIR Veracity and Study of Analysis Error Source[J].Spectroscopy and Spectral Analysis,2007,27(9):1751-1753.
Authors:LI Jun-hui  QIN Xi-yun  ZHANG Wen-juan  CAI Gui-min  YANG Yu-hong  ZHAO Long-lian  CHANG Zhi-qiang  ZHAO Li-li  ZHANG Lu-da
Institution:1. College of Information and Electrical Engineering,China Agricultural University, Beijing 100094, China2. Yunnan Tobacco Science Research Institute, Yuxi 653100, China3. College of Science, China Agricultural University, Beijing 100094, China
Abstract:The influence of sample test conditions on the NIR veracity was studied with homemade grating diffuse NIR instrument using Yunnan flue-cured tobacco.Deducing analysis error was achieved by model self-emendation when a global NIR model was set up.Without regarding the influence of loading samples and test conditions,the test repetition error,re-loading error and samples tightness error,which were brought by instrument S/N,accounted for 50%,30% and 20% of the total error,respectively.Depressing sample could reduce errors brought by sample tightness.Changes in test conditions could bring more analysis error,which was larger than the total of repetition error.These results theoretically explain the influence of sample test conditions on the NIR analysis veracity,which can provide basic theory data for farther improvement of homemade instrument and offer a new idea for resolving this problem.
Keywords:NIR  Flue-cured tobacco  Loading sample  Test condition  Error source