首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Baseline optimization of SQUID gradiometer for magnetocardiography
Authors:Li Hua  Zhang Shu-Lin  Qiu Yang  Zhang Yong-Sheng  Zhang Chao-Xiang  Kong Xiang-Yan  Xie Xiao-Ming
Institution:a State Key Laboratory of Functional Materials for Informatics, Shanghai Institute of Microsystem and Information Technology (SIMIT), Chinese Academy of Sciences (CAS), Shanghai 200050, China; b Joint Research Laboratory on Superconductivity and Bioelectronics, Collaboration Between CAS-Shanghai, Shanghai 200050, People''s Republic of China and FZJ, D-52425 Julich, Germany; c University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:SQUID gradiometer techniques are widely used in noise cancellation for biomagnetic measurements.An appropriate gradiometer baseline is very important for the biomagnetic detection with high performance.By placing several magnetometers at different heights along the vertical direction,we could simultaneously obtain the synthetic gradiometers with different baselines.By using the traditional signal-to-noise ratio(SNR) as a performance index,we successfully obtain an optimal baseline for the magnetocardiography(MCG) measurement in a magnetically shielded room(MSR).Finally,we obtain an optimal baseline of 7 cm and use it for the practical MCG measurement in our MSR.The SNR about 38 dB is obtained in the recorded MCG signal.
Keywords:SQUID  gradiometer  baseline optimization  magnetocardiography
本文献已被 CNKI 等数据库收录!
点击此处可从《中国物理 B》浏览原始摘要信息