首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Factors in the design of ultrasonic probes
Authors:WMR Smith  AO Awojobi
Institution:1. National Engineering Laboratory, East Kilbridge, Glasgow G75 0QU, UK;2. PO Box 1, Oshodi, Nigeria
Abstract:This paper attempts to demonstrate the comprehensive way in which the Unit Impulse Response technique can be used to evaluate and calibrate thickness mode piezoelectric transducers and the ultrasonic probes made from these transducers. Two commonly used piezoelectric materials were compared in the form of batches of disc transducers at several different frequencies and diameters.At higher system gains spurious signals associated with transverse mode resonances seriously reduced the discrimination of the probe. These spurious effects were fully investigated and explained.The effects of backing; quarter-wavelength matching into water; and electrical tuning, were measured and explained.Other methods of measuring the transducer parameters based on frequency response analysis were considered.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏