激光多模式毛细管电泳检测器的光学设计及优化
引用本文:余振宇,熊博,王玮,周平,胡继明.激光多模式毛细管电泳检测器的光学设计及优化[J].光子学报,2006,35(8):1248-1253.
作者姓名:余振宇  熊博  王玮  周平  胡继明
作者单位:1. 武汉大学电子信息学院测控技术与仪器系,武汉,430072
2. 武汉大学分析生物医学研究所,武汉,430072
3. 武汉大学电子信息学院测控技术与仪器系,武汉,430072;武汉大学分析生物医学研究所,武汉,430072
摘    要:提出一种基于毛细管电泳的激光多模式检测器.为设计并优化其光路构型,进行了理论分析和数学推导,并采用Matlab仿真,得到光路中各参量优化取值范围.在热透镜通道,激发光束腰半径越小,则热透镜效应越强.而探测激光束腰半径、束腰与样品距离、样品与探测面距离三个参量综合决定检出信号强度;在回射干涉通道,聚焦透镜焦距应较短,它与毛细管距离对检测影响很大,而与激光器、与探测面的两距离对检测影响相对较小;热透镜效应对回射干涉检测影响不明显,如要严格消除,可将两通道错开一适当距离.

关 键 词:分析仪器  光学设计  激光热透镜  激光回射干涉  激光诱导荧光
收稿时间:2005-12-03
修稿时间:2005年12月3日

The Optical Design and Optimization of Laser Based Multimode Detector for Capillary Electrophoresis
Yu Zhenyu,Xiong Bo,Wang Wei,Zhou Ping,Hu Jiming.The Optical Design and Optimization of Laser Based Multimode Detector for Capillary Electrophoresis[J].Acta Photonica Sinica,2006,35(8):1248-1253.
Authors:Yu Zhenyu  Xiong Bo  Wang Wei  Zhou Ping  Hu Jiming
Institution:1.Measure and Control Technology and Instrument Department ,Electronic Information College, Wuhan University, Wuhan 430072;2 .Graduate School of Analysis of Biology and Medicine, Wuhan University, Wuhan 430072
Abstract:A laser-based multimode detector is introduced.To gain the optimal detection,the optical configuration is designed and optimized.Based on the theoretical analysis,mathematical deduction and Matlab simulation,the optimum numerical range of the parameters of the optical configuration are obtained.In the thermal lens detection mode,generally,smaller waist radius induces stronger the thermal lens effect,on the other hand,the waist radius of the probe beam,the distance between the waist and the sample and the distance between the waist and the detecting plane synthetically determine the intensity of the acquired signal.In the retroreflected detection mode,the focus length of the focusing lens has to be short enough,and the distance between the lens and the capillary,which has to be less,influences the detection a lot,while the lens has to be far enough both from the laser source and the detection plane.Finally,the thermal lens impacts the retroreflected interference unapparently.To eliminate the influence strictly,a proper little distance to shift the two detecting channels can be adopted.
Keywords:Analytical instrument  Optical design  Laser thermal lens  Laser retroreflected beam interference  Laser-induced fluorescence
本文献已被 CNKI 维普 万方数据 等数据库收录!