Geometric calibration for a SPECT systemdedicated to breast imaging
作者姓名:武丽伟  曹学香  王璐  黄先超  柴培  贠明凯  张玉包  张龙  单保慈  魏龙
作者单位:1. Key Laboratory of Nuclear Analysis Techniques, Institute of High Energy Physics,Chinese Academy of Sciences, Beijing 100049, China;;2. Graduate University of Chinese Academy of Sciences, Beijing 100049, China;;3. Beijing Engineering Research Center of Radiographic Techniques and Equipment, Beijing 100049, China
基金项目:Supported by National Natural Science Foundation of China (118110117465)
摘    要:Geometric calibration is critical to the accurate SPECT reconstruction. In this paper, a geometric calibration method was developed for a dedicated breast SPECT system with a tilted parallel beam (TPB) orbit. The acquisition geometry of the breast SPECT was firstly characterized. And then its projection model was established based on the acquisition geometry. Finally, the calibration results were obtained using a nonlinear optimization method that fitted the measured projections to the model. Monte Carlo data of the breast SPECT were used to verify the calibration method. Simulation results showed that the geometric parameters with reasonable accuracy could be obtained by the proposed method.

关 键 词:geometric  calibration  acquisition  geometry  tilted  parallel  beam  (TPB)  breast  SPECT  Monte  Carlo  simulation
收稿时间:2012-1-17
修稿时间:2012-2-9

Geometric calibration for a SPECT system dedicated to breast imaging
WU Li-Wei,CAO Xue-Xiang,WANG Lu,HUANG Xian-Chao,CHAI Pei,YUN Ming-Kai,ZHANG Yu-Bao,ZHANG Long,SHAN Bao-Ci,;,WEI Long.Geometric calibration for a SPECT systemdedicated to breast imaging[J].High Energy Physics and Nuclear Physics,2012,36(10):1019-1024.
Authors:WU Li-Wei  CAO Xue-Xiang  WANG Lu  HUANG Xian-Chao  CHAI Pei  YUN Ming-Kai  ZHANG Yu-Bao  ZHANG Long  SHAN Bao-Ci  ;  WEI Long
Institution:1,3 1 Key Laboratory of Nuclear Analysis Techniques, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China 2 Graduate University of Chinese Academy of Sciences, Beijing 100049, China 3 Beijing Engineering Research Center of Radiographic Techniques and Equipment, Beijing 100049, China
Abstract: