Preliminary study of depth of interaction measurement for a PET detector
Authors:LIAO Yan-Fei    ZHANG Zhi-Ming  LI Dao-Wu  SHUAI Lei  SHAN Bao-Ci  HUANG Xian-Chao  LIU Jun-Hui  CHEN Yan  WANG Ying-Jie  LIU Shuang-Quan  WANG Pei-Lin  WEI Shu-Jun  WEI Long  Key Laboratory of Nuclear Analysis Techniques
Institution:LIAO Yan-Fei 1,2,3 ZHANG Zhi-Ming 1,2 LI Dao-Wu 1,2 SHUAI Lei 1,2 SHAN Bao-Ci1,2 HUANG Xian-Chao 1,3 LIU Jun-Hui 1,3 CHEN Yan 1,3 WANG Ying-Jie 1,3 LIU Shuang-Quan 1,3 WANG Pei-Lin 1,3 WEI Shu-Jun1,2 WEI Long1,1) 1 Key Laboratory of Nuclear Analysis Techniques,Institute of High Energy Physics,Chinese Academy of Sciences,Beijing 100049,China 2 Beijing Engineering Research Center of Radiographic Techniques and Equipment,China 3 Graduate University of Chinese Academy of Scien...
Abstract:In this work we studied the feasibility of detecting the depth of interaction (DOI) with two layers of crystal arrays of LYSO and BGO scintillators coupled to a position-sensitive photomultiplier tube (PS-PMT) R8900-C12. A front-end electronics was designed, with which we got different pulse shapes for different crystals to obtain depth information. With the double integration method, we got the DOI histogram of a divided integration ratio of two crystals as the standard to determine the layer-of-interactio...
Keywords:PET  detector  DOI  spatial resolution  
本文献已被 CNKI 维普 等数据库收录!
点击此处可从《中国物理 C》浏览原始摘要信息