Monitoring the energy variation of an electron linac using a Cerenkov detector
作者姓名:李树伟  王义  李金  李元景  康克军  林德旭  张清军
作者单位:Department of Engineering Physics, Tsinghua University, Beijing 100084, China; Key Laboratory of Particle & Radiation Imaging (Tsinghua University), Ministry of Education, Beijing 100084, China; Nuctech Company Limited, Beijing 100084, China
摘    要:A new method to monitor the energy variation of a multi-energy electron linac by combining a Cerenkov detector and a CsI(Tl) detector is reported. The signals in the Cerenkov detector show an appreciable but different dependence on the energy of the electron linac from the traditional CsI(Tl) detector due to the particular response of the former to charged electrons with high velocity above threshold. The method is more convenient than the HVL (half-value layer) method which is commonly employed to calibrate the energy of an electron linac for real time monitoring. The preliminary validity of the method is verified in a dual-energy electron linac with 6 MeV and 3 MeV gears. Moreover, the method combining the Cerenkov detector and the CsI(Tl) detector is applicable to probe the X-ray spectrum hardened by the inspected material and may serve as a novel tool for material discrimination with effective atomic number in radiation imaging.

关 键 词:电子直线加速器  能量变化  探测器  监测  碘化铯  信号显示  实时监控  原子序数
收稿时间:2009-2-13
修稿时间:2009-4-1

Monitoring the energy variation of an electron linac using a Cerenkov detector
Institution:Tsinghua University
Abstract:A new method to monitor the energy variation of a multi-energy electron linac by combining a Cerenkov detector and a CsI(Tl) detector is reported. The signals in the Cerenkov detector show an appreciable but different dependence on the energy of the electron linac from the traditional CsI(Tl) detector due to the particular response of the former to charged electrons with high velocity above threshold. The method is more convenient than the HVL (half-value layer) method which is commonly employed to calibrat...
Keywords:electron linac  Cerenkov detector  Geant4  