首页 | 本学科首页   官方微博 | 高级检索  
     检索      

Digital discrimination of neutrons and γ rays with organic scintillation detectors in an 8-bit sampling system using frequency gradient analysis
引用本文:杨俊,罗晓亮,刘国福,林存宝,王艳玲,胡青青,彭进先.Digital discrimination of neutrons and γ rays with organic scintillation detectors in an 8-bit sampling system using frequency gradient analysis[J].中国物理 C,2012,36(6):544-551.
作者姓名:杨俊  罗晓亮  刘国福  林存宝  王艳玲  胡青青  彭进先
作者单位:Department of Instrument Science and Technology, National University of Defense Technology, Changsha 410073, China
基金项目:Supported by National Natural Science Foundation of China (A050508/11175254)
摘    要:The feasibility of using frequency gradient analysis (FGA), a digital method based on Fourier transform, to discriminate neutrons and γ rays in the environment of an 8-bit sampling system has been investigated. The performances of most pulse shape discrimination methods in a scintillation detection system using the time-domain features of the photomultiplier tube anode signal will be lower or non-effective in this low resolution sampling system. However, the FGA method using the frequency-domain features of the anode signal exhibits a strong insensitivity to noise and can be used to discriminate neutrons and γ rays in the above sampling system. A detailed study of the quality of the FGA method in BC501A liquid scintillators is presented using a 5 G samples/s 8-bit oscilloscope and a 14.1 MeV neutron generator. A comparison of the discrimination results of the time-of-flight and conventional charge comparison (CC) methods proves the applicability of this technique. Moreover, FGA has the potential to be implemented in current embedded electronics systems to provide real-time discrimination in standalone instruments.

关 键 词:n-γ  discrimination  liquid  scintillator  FGA  8-bit  sampling  system
收稿时间:2011-10-8

Digital discrimination of neutrons and γ rays with organic scintillation detectors in an 8-bit sampling system using frequency gradient analysis
YANG Jun LUO Xiao-Liang,LIU Guo-Fu LIN Cun-Bao WANG Yan-Ling HU Qing-Qing PENG Jin-Xian.Digital discrimination of neutrons and γ rays with organic scintillation detectors in an 8-bit sampling system using frequency gradient analysis[J].High Energy Physics and Nuclear Physics,2012,36(6):544-551.
Authors:YANG Jun LUO Xiao-Liang  LIU Guo-Fu LIN Cun-Bao WANG Yan-Ling HU Qing-Qing PENG Jin-Xian
Institution:Department of Instrument Science and Technology,National University of Defense Technology,Changsha 410073,China
Abstract:The feasibility of using frequency gradient analysis(FGA),a digital method based on Fourier transform,to discriminate neutrons and γ rays in the environment of an 8-bit sampling system has been investigated.The performances of most pulse shape discrimination methods in a scintillation detection system using the time-domain features of the photomultiplier tube anode signal will be lower or non-effective in this low resolution sampling system.However,the FGA method using the frequency-domain features of the anode signal exhibits a strong insensitivity to noise and can be used to discriminate neutrons and γ rays in the above sampling system.A detailed study of the quality of the FGA method in BC501A liquid scintillators is presente dusing a 5 G samples/s 8-bit oscilloscope and a 14.1 MeV neutron generator.A comparison of the discrimination results of the time-of-?ight and conventional charge comparison(CC) methods proves the applicability of this technique.Moreover,FGA has the potential to be implemented in current embedded electronics systems to provide real-time discrimination in standalone instruments.
Keywords:n-γ discrimination  liquid scintillator  FGA  8-bit sampling system
本文献已被 CNKI 等数据库收录!
点击此处可从《中国物理 C》浏览原始摘要信息
点击此处可从《中国物理 C》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号