Determination of the time resolution for neutron scintillation detectors by multi-coincidence measurement
Authors:LI Yong-Ming    RUAN Xi-Chao  ZHOU Bin MA Zhong-Yuan State Key Laboratory of Nuclear Physics  Technology  Peking University  Beijing  China Science  Technology on Nuclear Data Laboratory  China Institute of Atomic Energy  Beijing  China
Institution:LI Yong-Ming 1,2,1) RUAN Xi-Chao 2,2) ZHOU Bin 2 MA Zhong-Yuan 2 1 State Key Laboratory of Nuclear Physics and Technology,Peking University,Beijing 100871,China 2 Science and Technology on Nuclear Data Laboratory,China Institute of Atomic Energy,Beijing 102413,China
Abstract:Based on the multi-coincidence measurement, the time resolution of three liquid scintillation detectors (BC501A) were determined strictly by solving the coincidence equations, where the influence from electronics estimated by self coincidence measurement as well as the background had been considered. The result of this work agreed well with the result that was deduced from the traditional method, and it will be helpful to analyze the energy resolution of neutron time of flight spectra measured by using such...
Keywords:BC501A detectors  multi-coincidence measurement  time resolution  