Abstract:The use of Bi-2223 high temperature superconducting (HTS) tape as a material for gradient coils in MRI is evaluated in this paper. Bi-2223 tapes have a very high critical current and a very low power loss. A HTS tape gradient coil is expected to provide much higher gradient strength and generate much lower heating than a copper coil. Measurements of the AC power loss of Bi-2223 tapes at typical operating frequencies for gradient coils are presented. The degradation of the critical current and its effect on the increase of AC power loss is analyzed. Practical technical issues such as resistance, gradient strength and mechanical performance are also discussed. A prototype Bi-2223 HTS tape gradient coil is evaluated to verify the concept.