Intensity of an oxygen saturation image improved using scanning NIR-LED light irradiation |
| |
Authors: | Hsin-Yi Tsai Ching-Ching Yang Min-Wei Hung Kuo-Cheng Huang |
| |
Institution: | 1. Instrument Technology Research Center, National Applied Research Laboratories, Hsinchu, 30076, Taiwan
|
| |
Abstract: | A near-infrared (940 nm) image of skin tissue can be substantially enhanced using optical scanning technology (OST) that can reduce the scattering effect and remove the wrinkles or fine hair in the image. Using OST to create the skin tissue image ensured that the contour of blood vessels became more clear than that obtained from the divergent light source. In addition, the oxygen saturation of the blood vessels and tissues obtained using OST were approximately 98.65 and 93.17%, respectively, exhibiting an increase of more than 20% compared with that obtained using divergent light, and approached the value when measured using commercial pulse oximetry. Because the proposed method has a deep light penetration depth and high image intensity and resolution for oxygen saturation analysis, it is highly appropriate to be applied to future studies on diabetes and cardiovascular diseases. |
| |