首页 | 本学科首页   官方微博 | 高级检索  
     检索      


The effects of hydroxyapatite coatings on stress distribution near the dental implant-bone interface
Authors:W Jiang  XH Shi  W Zou  JM Luo  ZW Gu
Institution:a Engineering Research Center for Biomaterials, Sichuan University, Chengdu, Sichuan 610064, China
b Center for Certification of Drug, State Food and Drug Administration, Beijing 100061, China
Abstract:The effects of different thickness of hydroxyapatite (HA) coatings on bone stress distribution near the dental implant-bone interface are very important factors for the HA-coated dental implant design and clinical application. By means of finite element analysis (FEA), the bone stress distributions near the dental implant coated with different thicknesses from 0 to 200 μm were calculated and analyzed under the 200 N chewing load. In all cases, the maximal von Mises stresses in the bone are at the positions near the neck of dental implant on the lingual side, and decrease with the increase of the HA coatings thickness. The HA coatings weaken the stress concentration and improve the biomechanical property in the bone, however, in HA coatings thickness range of 60-120 μm, the distinctions of that benefit are not obvious. In addition, considering the technical reason of HA coatings, we conclude that thickness of HA coatings range from 60 to 120 μm would be the better choice for clinical application.
Keywords:68  35  Gy  46  70  Lk
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号