首页 | 本学科首页   官方微博 | 高级检索  
     检索      

石墨炉原子吸收法测定人红细胞膜结合硒
引用本文:徐德选,杨根元.石墨炉原子吸收法测定人红细胞膜结合硒[J].光谱实验室,1996,13(2):20-23.
作者姓名:徐德选  杨根元
作者单位:镇江医学院分析化学教研室
摘    要:本文采用镍-硝酸镁作基体改进剂,用石墨炉原子地测定人红细胞膜结合硒,并对该方法进行了研究探索,在选定的实验条件下,该方法特征质量为16.0pg/0.0044A.S,检出限为7.4pg,回收率为92%-108%,相对标准偏差为8.9%,10种共存元素含量在正常生理浓度范围内不干扰测定,经过样品测试分析,证明本方法准确,快速,满足分析要求。

关 键 词:原子吸收法  红细胞膜  

Determination of Binding Selenium in Human Red Erythrocyte Membrane by Graphite Furnace Atomic Absorption Spectrometry
Xu Dexuan, Yang Genyuan.Determination of Binding Selenium in Human Red Erythrocyte Membrane by Graphite Furnace Atomic Absorption Spectrometry[J].Chinese Journal of Spectroscopy Laboratory,1996,13(2):20-23.
Authors:Xu Dexuan  Yang Genyuan
Abstract:A method for determination of binding selenium in human red erythrocyte membrane with Ni and Mg (NO_3)_2 as matrix modifier in graphite furnace AAS is described. Under the optimum condition,the characteristic mass m is 16. 0pg/0. 0044A. S,the detection limit is 7.4pg,the recovery is 92%-108 % ,the ralative standard deviation is less than 8. 9%.10 elements do not interfere with the determination of Se. The method had been used to analyse practical samples ,satisfactory results are obtained. It is simple and rapid.
Keywords:Graphite Furnace Atomic Absorption Spectrometry  Red Erythrocyte Membrane  Selenium    